The network plug-in isolator 100G provides a safe plug-in of medical devices 
and personal computers within the patient's vicinity into an IEEE 803.2 (10/100- 
BaseT,Twisted-Pair) computerNetwork. Workstations are allowed to operate within 
the patient's, only if they are designed in compliance with IEC 60601-1 as a 
medical device or if they are fitted with an IEC 60601-1 compliant electrical 
protection system. 
All Inputs and Outputs of a medical device to other non medical devices or 
Networks outside the surgery environment must comply with IEC-60601-1-1 
standard. This galvanic isolation with sufficiently high dielectric strength is 
required to meet the IEC-60601-1-1 standard. The Network plug-in isolator 100G 
may also be vicinity used in non medical applications. It provides galvanic 
isolation for any equipment connected via Ethernet ports to prevent ground 
loops.